Why the study?
Does the radioactive microsphere method accurately measure cardiac output compared to the indocyanine green dye dilution method in a rabbit model?
Does the radioactive microsphere method accurately measure cardiac output compared to the indocyanine green dye dilution method in a rabbit model?
The radioactive microsphere method provides a simple and accurate way to measure regional blood flow and cardiac output in small laboratory animals.
Supports microsphere accuracy for cardiac output in rabbits; extends validation for regional flow studies in small animals.
The determination of regional blood flow and cardiac output with radioactively labeled microspheres and a reference arterial sample has been assessed in the rabbit. Paired determinations of cardiac output with the microsphere and indocyanine green dye dilution methods correlated well (F=236; p<< 0.001; N=98). A simple method for administering microspheres with a 1.5% accuracy has been developed. Paired samples drawn simultaneously revealed mixing and streaming errors of 5.3%. Regional flows to several organs determined by the method were in accord with published data. The advantages of microsphere-reference sample method include: simplicity, accuracy, the provision of regional flows without simultaneously determining cardiac output, and applicability to small laboratory animals and low flow states.
No takes yet. Share an insight, caveat, or question.
Bartrum et al. (1974) studied this question.