Key result
The blanked flutter search algorithm in a dual-chamber pacemaker created a short-long-short cycle length alternation during sinus tachycardia, triggering an episode of ventricular tachycardia.
Case Report (n=1)
The blanked flutter search algorithm in pacemakers can inadvertently induce short-long-short cycle length alternations, potentially triggering ventricular tachycardia in susceptible patients.
Answers are grounded in this paper and cite their sources.
Alerts to proarrhythmic risk of blanked flutter search algorithms during sinus tachycardia; leaves open need for algorithm modifications or targeted monitoring.
Blanked flutter search (BFS) is an algorithm designed to unmask persistent 2:1 atrial flutters in which every other atrial event falls within atrial blanking period, concealing the arrhythmia. When the sensed atrial cycle length (CL) is less than: (i) twice the sum of atrioventricular interval (AVI) and post-ventricular atrial blanking period (PVAB); and (ii) twice the mode switch detection interval (MSDI) for eight consecutive ventricular cycles, the post-ventricular atrial refractory period (PVARP) is then extended for one cycle to search for a concealed atrial tachyarrhythmia. A patient with hypertrophic cardiomyopathy, left ventricular ejection fraction 72%, and high-degree atrioventricular block received a rate-adaptive dual chamber pacemaker (Medtronic EnPulse, MSDI-400ms, and PVAB-180ms). An episode of sinus tachycardia (CL: 523–539 ms) with ventricular tracking (AVI-93ms) fulfilled the criteria to initiate BFS. Post-ventricular atrial refractory period was extended for one cycle and a sinus impulse (*) was not tracked and conducted intrinsically (interval: 289 ms, dotted line). The next sinus impulse (**) was within PVARP and conducted. A short–long–short CL alternation was created and triggered ventricular tachycardia (arrow). The patient felt chest tightness and dyspneic and the VT terminated spontaneously 23 min later. Ventricular burst pacing in electrophysiologic laboratory induced a VT with same CL. An implantable cardioverter-defibrillator was implanted. Conflict of interest: none declared.
No takes yet. Share an insight, caveat, or question.
Chang et al. (2011) conducted a case report in Hypertrophic cardiomyopathy and high-degree atrioventricular block (n=1). Blanked flutter search (BFS) algorithm was evaluated on Ventricular tachycardia. The blanked flutter search algorithm in a dual-chamber pacemaker created a short-long-short cycle length alternation during sinus tachycardia, triggering an episode of ventricular tachycardia.