Why the study?
To measure high-sensitive troponin inter-identical-instrument bias to determine if it compromises the interpretation of absolute delta values for rapid 0/1-0/3 hour algorithms.
Does inter-identical-instrument bias compromise the interpretation of absolute high-sensitive troponin delta values around the 99th percentile upper reference limit?
Does inter-identical-instrument bias compromise the interpretation of absolute high-sensitive troponin delta values around the 99th percentile upper reference limit?
Inter-identical-instrument bias for high-sensitive troponin assays must be considered when defining the minimum absolute delta for rapid 0/1-0/3 hour diagnostic algorithms.
Supports consistent hs-troponin deltas across identical instruments; leaves open instrument-specific thresholds for rapid diagnostic algorithms.
Aim of the study: To measure high-sensitive troponin (Access hsTnI) interidentical- instrument bias (DxI800 Beckman Coulter) in order to understand if it can compromise the interpretation of absolute delta value for rapid algorithms 0/1-0/3 hours. Materials and methods: One hundred fifty-nine lithium/heparin plasma samples were processed sequentially on three DxI800 (DxI1, DxI2, DxI3). The results given by the three instruments were analyzed as followed: DxI1 vs. DxI2, DxI1 vs. DxI3, DxI2 vs. DxI3. Statistical analysis was done using the Passing-Bablok regression, Bland-Altman test, and Cohen’s Kappa statistic. Results: PB regression did not show any significant deviation from linearity and no proportional nor constant differences were observed among instruments. Moreover, the mean absolute bias, even though among the three instruments the lowest 95%CI lower limit was -3.75 and the highest 95%CI upper limit was 3.92 ng/L, was within the acceptance limits (all results<reference change value). The concordance between each couple of instruments was mostly strong. Conclusion: Our data suggest that inter-identical-instrument bias needs to be considered before evaluating the clinical diagnostic accuracy of one absolute delta with respect to another, in order to define the minimum absolute delta that the laboratory can guarantee to the clinicians.
No takes yet. Share an insight, caveat, or question.
Stenner et al. (2019) studied this question.