Answers are grounded in this paper and cite their sources.
A high-pressure absorption cell, designed to give sample pressure up to 50 kb at 90°K, and allowing sufficient transmission of radiation for infra-red absorption spectra to be recorded with a resolution of 1 cm -1 , is described. This cell has been working for the past four years to the above specification in the near infra-red region of the spectrum. It has also been used in other spectral regions (400-50 000 cm -1 ), and at other temperatures (77-400°K), and has permitted good spectra of slightly lower resolution to be obtained under these less favourable conditions.
No takes yet. Share an insight, caveat, or question.
W.F. Sherman (1966) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: