Key result
Standardized cardiac T2 mapping on identical 1.5 T CMR systems demonstrated high inter-scanner reproducibility with an overall mean myocardial T2 value of 45.7 ± 3.4 ms, despite significant regional variations.
Why the study?
Interpretation of cardiac T2 values remains difficult due to limited comparability across hardware and software systems and the lack of validated measurement recommendations for slice number and orientation.
Cross-Sectional (n=30)
Blinded observers
Randomized sequence of scanners
No
Standardized cardiac T2 mapping is highly reproducible across identical CMR systems, but significant physiological variations exist between myocardial segments, regions, and axes.
Supports standardized multi-center T2 mapping; leaves open need for segment-specific reference values.
BACKGROUND AND PURPOSE: Interpretation of T2 values remains difficult due to limited comparability across hardware and software systems and the lack of validated measurement recommendations for the number and orientation of mandatory slices. Our aims were to provide a standardized comparison of intra- and inter-individual T2 values in the short and long axes and to investigate inter-scanner reproducibility. METHOD AND MATERIALS: Ninety cardiovascular magnetic resonance (CMR) studies in 30 healthy subjects were performed with three identical 1.5 T CMR scanners (same hardware and software) using a balanced steady-state free precession (bSSFP) gradient echo sequence in three short axis (SAx) and three long axis (LAx) views. A commercially available T2 mapping software package of the latest generation with automatic in-line motion correction was used for acquisition. Regions of interest were manually drawn in each of the 16 myocardial segments according to the American Heart Association (AHA) model in three SAx and three LAx acquisitions. Analysis of inter-scanner, inter-segmental, intra-segmental, inter-regional and inter-level differences was performed. RESULTS: Inter-scanner reproducibility was high and the mean myocardial T2 value for all evaluated segments was 45.7 ± 3.4 ms. Significant inter-segmental variations of mean T2 values were found. Mean intra-segmental T2 values were comparable between LAx and SAx acquisitions in 72%. Significantly higher T2 values were found in apical segments and a significant disparity among different regions was found for SAx and LAx orientations. CONCLUSION: Standardized cardiac T2 mapping is highly reproducible on identical CMR systems. T2 values vary significantly between single heart segments, regions, levels, and axes in young, healthy subjects.
No takes yet. Share an insight, caveat, or question.
Wuest et al. (2020) conducted a cross-sectional in Healthy subjects (n=30). Standardized cardiac T2 mapping was evaluated on Mean myocardial T2 value and inter-scanner reproducibility. Standardized cardiac T2 mapping on identical 1.5 T CMR systems demonstrated high inter-scanner reproducibility with an overall mean myocardial T2 value of 45.7 ± 3.4 ms, despite significant regional variations.