Key result
Bipolar pacing vectors required lower capture threshold energy compared to pseudo-unipolar vectors at 15 months follow-up.
Why the study?
To monitor capture threshold changes over time for endovascular left ventricular pacing leads, compare pacing configurations, and assess the effect of steroid elution.
Does bipolar pacing compared to pseudo-unipolar pacing reduce capture threshold energy in patients receiving left ventricular leads for resynchronization therapy?
Observational (n=202)
No
Does bipolar pacing compared to pseudo-unipolar pacing reduce capture threshold energy in patients receiving left ventricular leads for resynchronization therapy?
Bipolar pacing vectors and steroid-eluting electrodes significantly reduce capture threshold energy requirements in left ventricular leads, which may improve device battery longevity.
Answers are grounded in this paper and cite their sources.
Bipolar steroid-eluting LV leads may favor lower energy use; leaves open effects on CRT outcomes and battery longevity.
Introduction The aim of this paper is to first monitor the changes in the capture threshold of endovascularly placed leads for left ventricle pacing, second to compare the pacing configurations, and third to verify the effect of Steroid elution for endovascular leads. Sample and Method The study included 202 consecutive single centre patients implanted with the Quartet™ lead (St. Jude Medical). The capture threshold and related lead parameters were tested during implantation, on the day of the patient’s discharge, and 3, 9, and 15 months after implantation. The electrical energy corresponding to the threshold values for inducing ventricular contraction was recorded for subgroups of patients with bipolar and pseudo-unipolar pacing vectors and electrodes equipped with and without a slow-eluting steroids. The best setting for the resynchronization effect was generally chosen. Capture threshold was taken as a selection criterion only if there were multiple options with (expected) similar resynchronization effect. Results and Discussion The measurements showed that the ratio of threshold energies of UNI vs. BI was 5× higher ( p < 0.001) at implantation. At the end of the follow-up, it dropped to 2.6 ( p = 0.012). The steroid effect in BI vectors was caused by a double capture threshold in the NSE group compared to the SE group ( p < 0.001), increased by approximately 2.5 times ( p < 0.001). The study concludes that after a larger initial increase in the capture threshold, the leads showed a gradual increase in the entire set. As a result, the bipolar threshold energies increase, and the pseudo-unipolar energies decrease. Since bipolar vectors require a significantly lower pacing energy, battery life of the implanted device would improve. When evaluating the steroid elution of bipolar vectors, we observe a significant positive effect of a gradual increase of the threshold energy.
No takes yet. Share an insight, caveat, or question.
Střítecký et al. (2023) conducted an observational in Cardiomyopathy indicated for resynchronization therapy (n=202). Bipolar pacing vector vs. Pseudo-unipolar pacing vector was evaluated on Capture threshold energy at 15 months (p=0.012). Bipolar pacing vectors required lower capture threshold energy compared to pseudo-unipolar vectors at 15 months follow-up.