Key result
Straight and spiral quadripolar leads achieve target proximal capture in ~82% of patients, favoring unipolar vectors.
Why the study?
Various lead designs exist for different coronary sinus anatomies, but comparative data on electrical parameters across straight and spiral designs, pacing vectors, and over time were needed.
Does lead design (straight vs spiral) and pacing vector affect electrical parameters of quadripolar coronary sinus leads in patients receiving left ventricular leads?
Cohort (n=795)
Does lead design (straight vs spiral) and pacing vector affect electrical parameters of quadripolar coronary sinus leads in patients receiving left ventricular leads?
Straight and spiral quadripolar leads provide clinically acceptable capture thresholds from proximal electrodes in over 80% of patients, with unipolar vectors offering lower thresholds than bipolar vectors.
Answers are grounded in this paper and cite their sources.
Proximal capture feasible with either quadripolar design in most patients; extends vector data but leaves practice change open pending randomized trials.
BACKGROUND: Various lead designs have been developed to accommodate different coronary sinus anatomies. Our objectives were to compare electrical parameters of straight and spiral left ventricular leads, to evaluate capture thresholds and impedances using different pacing vectors, and to study evolution of thresholds over time. METHODS: The RALLY-X4 study enrolled patients implanted with a lead from the Acuity X4 family (straight, spiral short, or spiral long). Electrical parameters (including capture thresholds from all 17 vectors) were measured at baseline and follow-up. RESULTS: Data from 795 patients who were successfully implanted were analyzed. Straight and spiral leads had similar proportions of patients with thresholds <2.5 V/0.4 ms using the distal electrode (61-65% of patients) or from at least one of the proximal (E2-E4) electrodes (81-83% of patients). Unipolar vectors had significantly lower thresholds and impedances than bipolar vectors, with similar measurements compared to extended bipolar configurations. Capture thresholds increased with more proximal electrodes for all leads. Over a mean follow-up of 1 year, a slight decrease in capture thresholds was observed. CONCLUSION: Straight and spiral quadripolar leads allow to obtain clinically acceptable capture thresholds from at least one of the proximal electrodes in >80% of patients. Pacing vectors significantly affect electrical parameters, with higher thresholds in more proximal electrodes and lower thresholds with unipolar and extended bipolar configurations. Capture thresholds slightly decreased over a mean follow-up of 1 year.
No takes yet. Share an insight, caveat, or question.
Burri et al. (2019) conducted a cohort in Patients requiring left ventricular lead implantation (n=795). Straight and spiral quadripolar left ventricular leads (Acuity X4 family) vs. Different lead designs and pacing vectors compared to each other was evaluated on Electrical parameters including capture thresholds and impedances. Straight and spiral quadripolar leads achieved capture thresholds <2.5 V/0.4 ms from at least one proximal electrode in 81-83% of patients, with unipolar vectors yielding lower thresholds.