Key result
ACUITY X4 quadripolar left ventricular leads achieved targeted pacing thresholds ≤2.5 V in 94% of patients and had an overall complication-free rate of 98% at 6 months.
Why the study?
Does the ACUITY X4 quadripolar left ventricular lead provide low pacing capture thresholds and low complication rates in patients undergoing CRT-D implantation?
Does the ACUITY X4 quadripolar left ventricular lead provide low pacing capture thresholds and low complication rates in patients undergoing CRT-D implantation?
The ACUITY X4 quadripolar LV leads demonstrated low pacing thresholds and a high complication-free rate at 6 months in patients receiving a CRT-D.
Answers are grounded in this paper and cite their sources.
Novel quadripolar leads feasible for nonapical CRT-D pacing; hypothesis-generating, RCTs needed before practice change.
INTRODUCTION: The safety and efficacy of a novel family of quadripolar left ventricular (LV) pacing leads designed to pace from nonapical regions of the LV with low pacing capture thresholds was studied in patients undergoing implantation of a cardiac resynchronization therapy defibrillator (CRT-D). METHODS AND RESULTS: Patients receiving a CRT-D were implanted with 1 of 3 ACUITY X4 leads (Spiral Long, Spiral Short, or Straight), designed to address coronary venous anatomical variability. Electrical performance and LV lead related complications were evaluated 3 and 6 months post implantation, respectively. 764 patients (68 ± 11 years, 66% male) were enrolled; 738 (97%) successfully implanted with an ACUITY X4 lead (Spiral L, n = 239, 31%; Spiral S, n = 281, 37%; Straight, n = 218, 29%). A targeted threshold ≤2.5 V was achieved in 644 (94%) patients. The median threshold from the best proximal electrode was lower than the tip electrode (0.9 V [IQR 0.7, 1.3] vs. 1.3 V [IQR 0.7, 2.5], p< 0.001) on Spiral leads. Irrespective of lead implanted, one of the proximal electrodes was the programmed cathode in most patients. The overall LV complication-free rate was 98%. LV lead dislodgment occurred in 8 (1%) patients. PNS occurred in 58 (8%) patients, but only 3 (0.4%) patients required surgical intervention. CONCLUSION: The ACUITY X4 LV leads had low pacing thresholds particularly from proximal electrodes, a high incidence of pacing from the nondistal electrode, and low likelihood of dislodgment or PNS requiring surgical intervention. (ClinicalTrials.gov Identifier: NCT02071173).
No takes yet. Share an insight, caveat, or question.
Mittal et al. (2016) studied Cardiac resynchronization therapy defibrillator (CRT-D) indication (n=764). ACUITY X4 quadripolar left ventricular pacing leads was evaluated on Electrical performance and LV lead related complications at 3 and 6 months. ACUITY X4 quadripolar left ventricular leads achieved targeted pacing thresholds ≤2.5 V in 94% of patients and had an overall complication-free rate of 98% at 6 months.