Key result
A 3D quadripolar spiral lead achieved similar average pacing capture thresholds in wedged versus unwedged positions and avoided phrenic nerve stimulation.
Why the study?
Does a novel 3D quadripolar lead design achieve acceptable pacing capture thresholds and avoid phrenic nerve stimulation in patients undergoing cardiac resynchronization?
Does a novel 3D quadripolar lead design achieve acceptable pacing capture thresholds and avoid phrenic nerve stimulation in patients undergoing cardiac resynchronization?
A novel 3D quadripolar spiral lead design achieves acceptable pacing capture thresholds and avoids phrenic nerve stimulation without repositioning during cardiac resynchronization therapy implants.
Answers are grounded in this paper and cite their sources.
Novel 3D quadripolar leads met acute PCT and PNS goals in most patients; leaves open long-term stability and CRT outcome benefits.
AIMS: This multicenter acute clinical study was designed to verify novel three-dimensional (3D) quadripolar lead designs that can achieve ≤2.5 V average pacing capture threshold (PCT) not only at the apex, but also at the base of the left ventricle with phrenic nerve stimulation (PNS) avoidance for cardiac resynchronization. METHODS: During the implant procedure, up to two different left ventricular investigational leads were introduced and tested in the same target coronary vein based on the coronary sinus venogram in a wedged and unwedged position. Adverse events were collected in 30 days following the procedure. RESULTS: Eighty-seven leads were tested in 50 patients. When the best performing spiral electrode was chosen from each lead testing, the average of the best PCT on spiral in a wedged position was similar to the unwedged position (1.7 ± 1.5 V vs 1.9 ± 1.5 V, P = ns) and was similar to the wedged tip electrode average PCT (1.7 ± 1.5 V vs 1.6 ± 1.6 V, P = ns). In the majority of patients (89-96%), pacing was achievable in a mid-basal ventricular location without PNS. CONCLUSIONS: This acute study demonstrated that a 3D quadripolar spiral lead design can achieve acceptable PCTs and avoid PNS without repositioning the lead at implant in the vast majority of patients. It also demonstrated that this lead design can achieve mid-basal ventricular stimulation with low PCT and good acute stability.
No takes yet. Share an insight, caveat, or question.
Philippon et al. (2015) studied Cardiac resynchronization (n=50). Three-dimensional (3D) quadripolar spiral lead vs. Unwedged position and wedged tip electrode was evaluated on Average pacing capture threshold (PCT) (p=ns). A 3D quadripolar spiral lead achieved similar average pacing capture thresholds in wedged versus unwedged positions and avoided phrenic nerve stimulation.