Why the study?
A point-of-care cTnI immunoassay on the AIA-360 analyzer was available for rapid diagnosis of myocardial injury in critical care, but had not been validated in canine and feline serum.
Does the AIA-360 immunoassay provide accurate and precise measurement of cTnI in canine and feline serum compared to the Immulite 1000 reference method?
Does the AIA-360 immunoassay provide accurate and precise measurement of cTnI in canine and feline serum compared to the Immulite 1000 reference method?
The AIA-360 immunoassay for cTnI in dogs and cats shows adequate precision but requires analyzer-specific reference intervals due to significant bias compared to the reference method.
Background Cardiac troponin I (cTnI) is a sensitive and specific biomarker for myocardial injury. Validation of point‐of‐care (POC) analyzers for cTnI measurement is valuable to the critical care setting, in which rapid results can facilitate prompt diagnoses. An immunoassay for detecting cTnI is available for the POC AIA‐360 analyzer (Tosoh Bioscience), but this has not been validated using canine and feline serum. Objectives The objectives were (a) to determine precision, accuracy, and linearity of cTnI measurement using the AIA‐360 immunoassay in pooled canine and feline samples, and (b) to compare results for individual canine and feline samples with those obtained using a reference chemiluminescence method (Immulite 1000, Siemens). Methods Intra‐ and inter‐assay repeatability was determined using pooled canine and feline samples, and the coefficient of variation (CV) was calculated for each. Pooled samples were also serially diluted to assess linearity. A modified spike and recovery analysis was performed by mixing pooled samples with different concentrations. Bland‐Altman and Deming regression analyses were used to determine bias for individual samples, and the total observed error (TE obs ) was calculated. Results Coefficient of variation values were well within the required maximum of 20%. Linearity was demonstrated over the range of samples tested, and the recovery study showed minimal proportional inaccuracies. Although the correlation between the analyzers was excellent, there was a large mean bias due to relative proportional bias. Total observed error consequently exceeded the total allowable error (TE A ). Conclusion Although, in most respects, the analyzer demonstrated adequate performance, pronounced bias contributed to the large TE obs , indicating a requirement for analyzer‐specific reference intervals.
No takes yet. Share an insight, caveat, or question.
Irvine et al. (2019) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: