Synapse
⌘+K
Synapse
PulseExploreClubsResearchersJournals
Instagram
HomeClubsExplore
January 1, 2006Journal of Veterinary Internal Medicine

Comparison of Canine Cardiac Troponin I Concentrations as Determined by 3 Analyzers

View Full Paper
Ask AI
Bookmark
Share

Key result

Canine cTnI values obtained from the Biosite, Stratus, and Access analyzers were closely correlated (r = 0.61 to 0.89) but had wide limits of agreement and cannot be directly compared.

Population

23 dogs with cardiac disease and reconstituted purified canine free cTnI diluted with canine plasma

Comparison

Measurement of cardiac troponin I using three… vs Comparison among the three analyzers

Design

Preclinical

Authors

DADarcy B. AdinUniversity of FloridaMOMark A. OyamaHeart Failure & TransplantMSMargaret M. SleeperFlorida College

Discussion

Loading...

Member takes

Implication

Cannot interchange canine cTnI values across analyzers in practice; leaves open assay standardization for veterinary research.

Study Design

Type

Observational (n=23)

Structured PICO

P
Population
23 dogs with cardiac disease evaluated for plasma cTnI concentrations using three different commercial analyzers.
E
Exposure
Measurement of cardiac troponin I (cTnI) using three commercial analyzers: Biosite Triage Meter, Dade-Behring Stratus, and Beckman-Coulter Access AccuTnI
C
Comparator
Comparison among the three analyzers
O
Outcome
Correlation and limits of agreement of cTnI concentrations among the three analyzerssurrogate

Main Result

Effect estimate: r = 0.61 to 0.89

Canine cTnI values obtained from different commercial analyzers are correlated but cannot be directly compared due to wide limits of agreement.

Limitations

  • In the absence of a gold standard none of the analyzers can be considered more correct than the others.

Cite This Study

Adin et al. (2006) conducted an observational in cardiac disease (n=23). Biosite Triage Meter, Dade-Behring Stratus, and Beckman-Coulter Access AccuTnI analyzers was evaluated on Correlation and agreement of canine plasma cTnI concentrations between 3 analyzers (r = 0.61 to 0.89). Canine cTnI values obtained from the Biosite, Stratus, and Access analyzers were closely correlated (r = 0.61 to 0.89) but had wide limits of agreement and cannot be directly compared.

synapsesocial.com/papers/6a8ec51ffb6e86649d28da51https://doi.org/10.1892/0891-6640(2006)20[1136:coccti]2.0.co;2
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Characterization of cardiac troponin subunit release into serum after acute myocardial infarction and comparison of assays for troponin T and I1998 · 397 citations
  2. 2Validation of the 99th Percentile Cutoff Independent of Assay Imprecision (CV) for Cardiac Troponin Monitoring for Ruling Out Myocardial Infarction2005 · 135 citations
  3. 3Cardiac Troponin-I Concentration in Dogs with Cardiac Disease2004 · 193 citations
  4. 4Cardiac troponin I in cats with hypertrophic cardiomyopathy2003 · 106 citations
  5. 5Cardiac Troponin I in the Normal Dog and Cat2001 · 130 citations