Why the study?
The Cobra PzF stent was designed to overcome shortfalls of BMS and DES in patients requiring short DAPT, but routine practice outcomes needed evaluation.
Does the Cobra PzF NanoCoated coronary stent provide safe and effective clinical outcomes in patients undergoing routine PCI who require short DAPT?
Does the Cobra PzF NanoCoated coronary stent provide safe and effective clinical outcomes in patients undergoing routine PCI who require short DAPT?
The Cobra PzF NanoCoated coronary stent demonstrates satisfactory 1-year clinical outcomes with low rates of MACE and stent thrombosis in a high-risk, real-world population requiring short DAPT.
Answers are grounded in this paper and cite their sources.
Supports short-DAPT feasibility with Cobra PzF in high-bleeding-risk patients; leaves open comparative efficacy versus contemporary DES.
BACKGROUND: The Cobra PzF coronary stent is cobalt chromium with flat thin struts, nano-coated with Polyzene-F that enhance rapid reendothelialization and reduce the risk of stent thrombosis and restenosis. It is designed to overcome shortfalls of BMS and DES in patients requiring short DAPT duration. AIMS: To report procedural and 1-year clinical outcomes following Cobra PzF stent implantation in routine practice PCI. METHODS: e-Cobra registry is a multicenter prospective study to evaluate Cobra PzF stent in routine practice in patients deemed appropriate for short DAPT after PCI. The primary endpoint was MACE rate at 12 months (Cardiac death, MI, TLR). The secondary endpoint was definite stent thrombosis at 12 months. RESULTS: Among 940 patients (72% men, 72.8 ± 13.4 years) with multiple co-morbidities, 47% had acute coronary syndromes, and 62% were defined as high bleeding risk. A total of 1,229 lesions were treated with 1,314 stents. 36% of patients had lesion type B2 or C classification. Angiographic success was achieved in all cases. One-year follow-up was available for 97% of patients. The primary endpoint occurred in 9.0% of patients, including cardiac death 3.7%, MI 4.8%, and TLR 4.3%. Definite stent thrombosis occurred in six out of 915 (0.7%). CONCLUSION: The Cobra PzF stent was safe and effective in routine practice patients and seems feasible in situations when short DAPT or Mono Antiplatelet Therapy (MAPT) is needed. One-year follow-up was associated with satisfactory clinical outcomes and validate previously reported data.
No takes yet. Share an insight, caveat, or question.
Maillard et al. (2020) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: