Why the study?
His-bundle pacing can be achieved via atrial-side or ventricular-side pacing, but comparisons of their pacing parameters and electrophysiological characteristics in bradycardia patients were needed.
Does ventricular-side His-bundle pacing improve pacing parameters compared to atrial-side His-bundle pacing in bradycardia patients?
Does ventricular-side His-bundle pacing improve pacing parameters compared to atrial-side His-bundle pacing in bradycardia patients?
Ventricular-side His-bundle pacing provides lower capture thresholds and higher R-wave amplitudes compared to atrial-side pacing, potentially improving battery longevity.
Ventricular-side His-bundle pacing may improve battery longevity; extends randomized evidence for optimized lead placement in bradycardia.
AIMS: His-bundle pacing (HBP) can be achieved in either atrial-side HBP (aHBP) or ventricular-side HBP (vHBP). The study compared the pacing parameters and electrophysiological characteristics between aHBP and vHBP in bradycardia patients. METHODS AND RESULTS: Fifty patients undergoing HBP implantation assisted by visualization of the tricuspid valvular annulus (TVA) were enrolled. The HBP lead position was identified by TVA angiography. Twenty-five patients were assigned to undergo aHBP and compared with 25 patients who underwent vHBP primarily in a prospective and randomized fashion. Pacing parameters and echocardiography were routinely assessed at implant and 3-month follow-up. His-bundle pacing was successfully performed in 45 patients (90% success rate with 44.4% aHBP and 55.6% vHBP). The capture threshold was lower in vHBP than aHBP at implant (vHBP: 1.1 ± 0.5 vs. aHBP: 1.4 ± 0.4 V/1.0 ms, P = 0.014) and 3-month follow-up (vHBP: 0.8 ± 0.4 vs. aHBP: 1.7 ± 0.8 V/0.4 ms, P < 0.001). The R-wave amplitude was higher in vHBP than in aHBP at implant (vHBP: 4.5 ± 1.4 vs. aHBP: 2.0 ± 0.8 mV, P < 0.001) and at 3-month follow-up (vHBP: 4.4 ± 1.5 vs. aHBP: 1.8 ± 0.7 mV, P < 0.001). No procedure-related complications and aggravation of tricuspid valve regurgitation were observed in most patients and echocardiographic assessment of cardiac function remained in the normal range in all patients during the follow-up. CONCLUSION: This study demonstrates that vHBP features a low and stable pacing capture threshold and high R-wave amplitude, suggesting better pacing mode management and battery longevity can be achieved by HBP in the ventricular side.
No takes yet. Share an insight, caveat, or question.
Hu et al. (2020) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: