A symmetrically opening slit mechanism, particularly suitable for infra-red spectrometers, is described. It employs parallel spring movements, operates for widths from 1 to 2 000 μ and is suitable for use with curved jaws. Its operating torque is about 50 g cm which, with its rapid response and freedom from backlash, makes it suitable for automatic control. The jaws remain parallel to within about 3 sec, and the mid-line of the slit does not drift more than about 2 μ. Hysteresis between the driving micrometer drum and the measured slit width amounts to 0.2-0.3 μ, and may be due to an oil film. The slit jaws can be set so close together as to polarize the transmitted light completely. The slit mechanism, with a suitable rotational movement, has been applied to a twin-beam infra-red spectrometer; it has been demonstrated that entrance and exit slits must be alined to within a few seconds if satisfactory balance of the twin beams is to be achieved. An appendix describes a new all-spring reversing mechanism which enables the slit width to be controlled to better than 0.05 μ, and which is apparently completely free from hysteresis.
No takes yet. Share an insight, caveat, or question.
R V Jones (1952) studied this question.