The instrument described may be used for the simultaneous measurement and time integration of potentials varying between 0 and 0.5 V. with an accuracy of 3%. The input impedance and grid current have values similar 10 10 ω and similar 10 -10 A. respectively, and the instrument can thus be used for measurement of currents down to 0.01 μA. full-scale deflexion, by suitable selection from the range of input resistors provided. Currents are read on a microammeter and time integration results are obtained as counts on a standard Post Office call register. Ordinary valves and components are used throughout. A detailed analysis of the Miller type of integrating circuit has shown that, when a larger value of input potential (say 10 V.) is available, an accuracy of within 0.2% is expected over most of the range. This has been verified experimentally.