The galvanometer amplifier described in this paper follows standard practice in that the beam of light reflected from the mirror of the primary galvanometer falls upon a pair of selenium rectifier photocells connected in series-opposition. The net current from the cells is indicated by a secondary galvanometer. Much improved stability and linearity are, however, secured by the use of a red-absorbing filter to eliminate photocell fatigue, and negative feedback from the photocell circuit to the primary galvanometer circuit. Sufficient gain to provide for the negative feedback is secured, without valve amplification, by using a sufficiently intense beam of light. The instrument is very simple to set up, robust galvanometers of standard pattern are used, and an unregulated 12 V. a.c. or d.c. supply for the galvanometer lamps is the only electric supply necessary. At a sensitivity (overall) of 15-20 m./μA. the instrument is quite stable in operation and zero repetition is within 10 -10 A. The simple theory is given and a number of uses for the instrument are suggested.