Key result
A newly developed real-time RT-PCR assay for EV-D68 demonstrated an analytic limit of detection of 4 copies per reaction and was more sensitive than commercially available respiratory panels.
Answers are grounded in this paper and cite their sources.
A newly developed RT-PCR assay provides highly sensitive and specific detection of EV-D68, which will be useful for identifying and studying outbreaks.
We have developed and evaluated a real-time reverse transcriptase PCR (RT-PCR) assay for the detection of human enterovirus D68 (EV-D68) in clinical specimens. This assay was developed in response to the unprecedented 2014 nationwide EV-D68 outbreak in the United States associated with severe respiratory illness. As part of our evaluation of the outbreak, we sequenced and published the genome sequence of the EV-D68 virus circulating in St. Louis, MO. This sequence, along with other GenBank sequences from past EV-D68 occurrences, was used to computationally select a region of EV-D68 appropriate for targeting in a strain-specific RT-PCR assay. The RT-PCR assay amplifies a segment of the VP1 gene, with an analytic limit of detection of 4 copies per reaction, and it was more sensitive than commercially available assays that detect enteroviruses and rhinoviruses without distinguishing between the two, including three multiplex respiratory panels approved for clinical use by the FDA. The assay did not detect any other enteroviruses or rhinoviruses tested and did detect divergent strains of EV-D68, including the first EV-D68 strain (Fermon) identified in California in 1962. This assay should be useful for identifying and studying current and future outbreaks of EV-D68 viruses.
No takes yet. Share an insight, caveat, or question.
Wylie et al. (2015) studied Human enterovirus D68 (EV-D68). EV-D68 real-time reverse transcriptase PCR (RT-PCR) assay vs. Commercially available assays was evaluated on Analytic limit of detection and sensitivity. A newly developed real-time RT-PCR assay for EV-D68 demonstrated an analytic limit of detection of 4 copies per reaction and was more sensitive than commercially available respiratory panels.
Synapse has enriched 10 related papers. Consider them for comparative context: