A low-power prescreening technique using impedance measurements can detect dangerous resonant interactions between MRI radiofrequency fields and conductive implants.
May support impedance prescreening for MRI implant safety; animal data leaves open clinical translation.
PURPOSE: The concept of a "radiofrequency safety prescreen" is investigated, wherein dangerous interactions between radiofrequency fields used in MRI, and conductive implants in patients are detected through impedance changes in the radiofrequency coil. THEORY: The behavior of coupled oscillators is reviewed, and the resulting, observable impedance changes are discussed. METHODS: A birdcage coil is loaded with a static head phantom and a wire phantom with a wire close to its resonant length, the shape, position, and orientation of which can be changed. Interactions are probed with a current sensor and network analyzer. RESULTS: Impedance spectra show dramatic, unmistakable splitting in cases of strong coupling, and strong correlation is observed between induced current and scattering parameters. CONCLUSIONS: The feasibility of a new, low-power prescreening technique has been demonstrated in a simple phantom experiment, which can unambiguously detect resonant interactions between an implanted wire and an imaging coil. A new technique has also been presented which can detect parallel transmit null modes for the wire.