Key result
Meeting ≥2 device-derived diagnostic criteria on discharge day increased the risk of 30-day heart failure readmission compared to meeting no criteria (28% vs 7%; HR 5.0; 95% CI 1.9-13.5; P=0.001).
Why the study?
Do device-derived diagnostic criteria evaluated on the day of discharge identify heart failure patients at increased risk for 30-day readmission?
Cohort (n=175)
Yes
Do device-derived diagnostic criteria evaluated on the day of discharge identify heart failure patients at increased risk for 30-day readmission?
Device-derived diagnostic criteria from CRT devices on the day of discharge can effectively stratify patients by their risk of 30-day heart failure readmission.
May aid discharge-day risk stratification in HF; hypothesis-generating and should not yet change practice.
AIMS: We hypothesized that diagnostic data in implantable devices evaluated on the day of discharge from a heart failure hospitalization (HFH) can identify patients at risk for HF readmission (HFR) within 30 days. METHODS AND RESULTS: In this retrospective analysis of four studies enrolling patients with CRT devices, we identified patients with a HFH, device data on the day of discharge, and 30-day post-discharge clinical follow-up. Four diagnostic criteria were evaluated on the discharge day: (i) intrathoracic impedance>8 Ω below reference impedance; (ii) AF burden>6 h; (iii) CRT pacing<90%; and (iv) night heart rate>80 b.p.m. Patients were considered to have higher risk for HFR if ≥2 criteria were met, average risk if 1 criterion was met, and lower risk if no criteria were met. A Cox proportional hazards model was used to compare the groups. The data cohort consisted of a total of 265 HFHs in 175 patients, of which 36 (14%) were followed by HFR. On the discharge day, ≥2 criteria were met in 43 (16% of 265 HFHs), only 1 criterion was met in 92 (35%), and none of the four criteria were met in 130 HFHs (49%); HFR rates were 28, 16, and 7%, respectively. HFH with ≥2 criteria met was five times more likely to have HFR compared with HFH with no criteria met (adjusted hazard ratio 5.0; 95% confidence interval 1.9–13.5, P=0.001). CONCLUSION: Device-derived diagnostic criteria evaluated on the day of discharge identified patients at significantly higher risk of HFR.
No takes yet. Share an insight, caveat, or question.
Small et al. (2013) conducted a cohort in Heart failure (n=175). Meeting ≥2 device-derived diagnostic criteria on discharge day vs. Meeting no criteria was evaluated on Heart failure readmission within 30 days (HR 5.0, 95% CI 1.9-13.5, p=0.001). Meeting ≥2 device-derived diagnostic criteria on discharge day increased the risk of 30-day heart failure readmission compared to meeting no criteria (28% vs 7%; HR 5.0; 95% CI 1.9-13.5; P=0.001).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: