Key result
Patients who developed decompensated heart failure had a greater range of daily shock impedance compared to those without decompensation.
Why the study?
Does daily shock impedance measured by an implantable cardioverter-defibrillator predict decompensated heart failure and correlate with BNP levels in patients with an ICD?
Observational (n=29)
No
Does daily shock impedance measured by an implantable cardioverter-defibrillator predict decompensated heart failure and correlate with BNP levels in patients with an ICD?
A sudden decrease in daily shock impedance measured by an ICD correlates with rising BNP levels and may serve as an early warning sign for decompensated heart failure.
Answers are grounded in this paper and cite their sources.
Greater impedance variability associated with decompensated HF; hypothesis-generating for ICD monitoring pending prospective validation.
BACKGROUND: Clinical data suggest that changes in intrathoracic impedance and fluid accumulation in the lung are inversely related. METHODS AND RESULTS: Daily shock impedance (S-IMP) was evaluated in 29 patients in whom a Ventak Prizm 2 was implanted (61+/-14 years old). The mean follow-up period was 45+/-18 weeks, during which 6 patients had episodes of decompensated heart failure (DHF group) and the others did not (NHF group). There was no significant difference between the DHF group and NHF group in the mean value of the S-IMP (46.4 +/-3.3 vs 45.4+/-5.4 Omega). The range of S-IMP in individual patients in the DHF group was significantly greater than that in the NHF group (13.8+/-0.38 vs 7.0+/-3.1 Omega, p<0.0001). Mean weekly change of S-IMP in individual patients in the DHF group was significantly greater than that in the NHF group (1.583+/-0.630 vs 1.092+/-0.361 Omega, p<0.0176). When the cut-off value was set at >1.242 Omega, sensitivity was 100% and specificity was 69.6% for a diagnosis of DHF. There was a significant negative correlation between the percent increases in brain natriuretic peptide (BNP) and S-IMP (correlation coefficient: -0.775 p<0.0001) in the DHF group. There was an inverse relation between BNP and S-IMP. CONCLUSIONS: Measurement of shock impedance may be useful in the management of congestive heart failure.
No takes yet. Share an insight, caveat, or question.
Matsushita et al. (2006) conducted an observational in Congestive heart failure (n=29). Decompensated heart failure (DHF) vs. Non-decompensated heart failure (NHF) was evaluated on Range of daily shock impedance (S-IMP) in individual patients (p=<0.0001). Patients who developed decompensated heart failure had a greater range of daily shock impedance compared to those without decompensation.
Synapse has enriched 10 related papers. Consider them for comparative context: