The EMIT serum digoxin assay can be successfully adapted for use with a mini-disc centrifugal analyzer, reducing required serum volume and measurement time.
Answers are grounded in this paper and cite their sources.
May enable smaller-volume, faster digoxin assays; leaves open clinical validation before routine use.
We have modified the EMIT (Syva) serum digoxin assay for use with a mini-disc centrifugal analyzer. A minicomputer is used to control data acquisition, calculations, and readout for 14 cuvettes simultaneously. Thus, it is not necessary for time each individual assay manually or to plot calibration curves on graph paper. We use incubation times and dilutions similar to those used for the manual procedure, but we have changed the buffer solution, decreased the measurment time, and decreased the required serum volume per assay fivefold. We evaluated day-to-day and within-run precision, using Syva calibrators and control sera. EMIT assay values for radioimmunoassay control sera correlate well with the supplier's stated values.