Key result
Modified EMIT 2000 digoxin assay matches the original assay and achieves RIA-level quantification limits.
Population
25 plasma specimens from patients receiving digoxin and 18 healthy volunteers
Comparison
Modified EMIT 2000 digoxin assay vs Original EMIT 2000 digoxin assay and literature…
Design
Other
Authors
Loading...
Supports modified digoxin assay for pharmacokinetic studies; leaves open routine clinical adoption pending further validation.
The modified EMIT 2000 digoxin assay provides sensitivity comparable to radioimmunoassay, making it suitable for pharmacokinetic interaction studies.
Radembino et al. (1999) studied Digoxin pharmacokinetics (n=43). Modified EMIT 2000 digoxin assay vs. Original EMIT 2000 digoxin assay was evaluated on Digoxin plasma concentrations. The modified EMIT 2000 digoxin assay demonstrated a lower limit of quantification of 0.08 ng/mL, similar to radioimmunoassay, and showed no significant difference compared to the original assay.