This case demonstrates a potential risk of inducing ventricular tachycardia solely by reducing the pacing rate during routine pacemaker follow-up.
Answers are grounded in this paper and cite their sources.
May suggest VT risk during routine pacemaker rate reduction; single case leaves open need for prospective validation.
Ventricular arrhythmias during a pacemaker follow-up have been previously reported, usually in conjunction with temporary asynchronous stimulation of a demand pacemaker through magnet application or by increased myocardial excitability, for example, following a myocardial infarction. The subject of this report, an 82-year-old pacemaker patient, had been VVI-paced without problems for the past 11 years. As an aid in determining the sensing threshold, the pacemaker lower rate was reduced from 70 to 40 beats/min. A ventricular tachycardia of 240 beats/min was induced, most likely following short-long cycles; syncope resulted. To our knowledge, this is the first report of induction of a ventricular tachycardia during pacemaker follow-up solely by reduction of pacing rate and not by asynchronous pacing. This case demonstrates an additional potential risk associated with pacemaker rate manipulation during pacemaker follow-up.
No takes yet. Share an insight, caveat, or question.
Alt et al. (1996) studied this question.