Key result
Non-invasive cardiovascular profiling using the HDI/Pulsewave CR-2000 system demonstrated short- and intermediate-term repeatability for arterial elasticity indices and agreed with invasive measures.
Why the study?
Is the HDI/Pulsewave CR-2000 Research CardioVascular Profiling System repeatable and does it agree with invasive hemodynamic measurements?
Observational (n=113)
Is the HDI/Pulsewave CR-2000 Research CardioVascular Profiling System repeatable and does it agree with invasive hemodynamic measurements?
The HDI/Pulsewave CR-2000 system provides repeatable non-invasive measurements of arterial elasticity that reasonably agree with invasive hemodynamic assessments.
Answers are grounded in this paper and cite their sources.
Cautions against routine clinical adoption; leaves open prospective validation of non-invasive arterial elasticity profiling.
OBJECTIVES: The aim of these studies was to assess repeatability of large (C1) and small (C2) arterial elasticity indices over various time intervals using the HDI/Pulsewave CR-2000 Research CardioVascular Profiling System (Hypertension Diagnostics, Inc., Eagan, Minnesota, USA). Non-invasive hemodynamic parameters using this device were compared to invasive measurements. METHODS: After a 5-min period of rest, 31 healthy hospital employees underwent cardiovascular profiling on two occasions within 1 h apart. Another 59 healthy hospital employees underwent cardiovascular profiling on two occasions, an average 52 days apart. An additional group of 23 patients underwent right and left heart catheterizations for routine clinical indications and hemodynamic assessment was performed invasively and non-invasively. RESULTS: For short-term repeatability, the mean difference of C1 was +0.25 +/- 2.83 ml/mmHg x 10 (P = NS) and C2 was -0.14 +/- 1.86 ml/mmHg x 100 (P = NS). For intermediate test repeatability, the mean difference of C1 was -0.415 +/- 2.97 ml/mmHg x 10 (P = NS) and C2 was -0.19 +/- 2.67 ml/mmHg x 100 (P = NS). In the invasive protocol, both aortic diastolic blood pressure (-4.74 +/- 9.7 mmHg) and systemic vascular resistance (-194 +/- 264 dyne x s x cm(-5)) were significantly lower invasively. CONCLUSIONS: Measurements with the HDI/Pulsewave CR-2000 Research CardioVascular Profiling System are repeatable over both a short and intermediate period of observation. Furthermore, non-invasive hemodynamic parameters reasonably agree with invasive measurements.
No takes yet. Share an insight, caveat, or question.
Prisant et al. (2002) conducted an observational in Healthy individuals and patients with clinical indications for catheterization (n=113). Non-invasive cardiovascular profiling (HDI/Pulsewave CR-2000) vs. Invasive hemodynamic measurements was evaluated on Repeatability of large (C1) and small (C2) arterial elasticity indices (p=NS). Non-invasive cardiovascular profiling using the HDI/Pulsewave CR-2000 system demonstrated short- and intermediate-term repeatability for arterial elasticity indices and agreed with invasive measures.