Key result
Arterial stiffness index measurement using the CardioVision MS-2000 demonstrated poor test-retest repeatability over 147 days, with an intraclass correlation coefficient of 0.25 (P=0.08).
Why the study?
Does the CardioVision MS-2000 provide repeatable measurements of arterial stiffness index in healthy adults?
Observational (n=47)
Does the CardioVision MS-2000 provide repeatable measurements of arterial stiffness index in healthy adults?
The CardioVision MS-2000 device demonstrates poor test-retest repeatability for measuring arterial stiffness index when using consecutive measurements.
Poor repeatability of CardioVision MS-2000 arterial stiffness index cautions against clinical use; leaves open need for prospective device validation.
BACKGROUND: Increased arterial stiffness is an early indicator of vascular disease. Several methods may be used to determine arterial stiffness. One method obtains an arterial stiffness index from the vascular dynamics of oscillometric-derived brachial artery pressure. METHODS: To determine the test-retest repeatability of the CardioVision MS-2000, 47 healthy hospital employees had five consecutive measurements of arterial stiffness index measured after a 5-10 min period of rest and then repeated after an average of 146.8 days. Their mean age was 37 years and 71% were women. RESULTS: The mean arterial stiffness index was 39.6+/-9.7 and 37.2+/-10.5 mmHg x10 (P=0.22) for the first and second time period, respectively. We computed an intraclass correlation coefficient of 0.31 and 0.33 for the first and second time periods, which is the measure of consistency or agreement of arterial stiffness index values within cases. The intraclass correlation coefficient for systolic blood pressure, diastolic blood pressure, heart rate and arterial stiffness index were 0.68 (P=0.0001), 0.70 (P=0.0001), 0.35 (P=0.02) and 0.25 (P=0.08), respectively. CONCLUSION: The results of this study suggest poor test-retest repeatability if consecutive measurements are used. The intraclass correlation coefficient, however, could be improved by eliminating the highest and lowest value from a set of measurements.
No takes yet. Share an insight, caveat, or question.
Sharma et al. (2005) conducted an observational in Healthy (n=47). CardioVision MS-2000 arterial stiffness index measurement was evaluated on Test-retest repeatability (intraclass correlation coefficient) of arterial stiffness index (ICC 0.25, p=0.08). Arterial stiffness index measurement using the CardioVision MS-2000 demonstrated poor test-retest repeatability over 147 days, with an intraclass correlation coefficient of 0.25 (P=0.08).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: