Why the study?
Does electrical muscle stimulation (EMS) therapy cause electromagnetic interference and inappropriate discharge in patients with different types of ICD sensing circuits?
Does electrical muscle stimulation (EMS) therapy cause electromagnetic interference and inappropriate discharge in patients with different types of ICD sensing circuits?
Electrical muscle stimulation can cause electromagnetic interference leading to inappropriate ICD discharge in patients with epicardial screw-in bipolar sensing leads, but may be safe in those with endocardial true bipolar sensing.
Answers are grounded in this paper and cite their sources.
EMS may cause inappropriate ICD discharge with epicardial leads; hypothesis-generating for endocardial true bipolar sensing safety.
This case report is about two patients with two different types of ICDs who underwent electrical muscle stimulation (EMS) therapy. In one patient with an ICD that has epicardial screw-in bipolar sensing leads, electromagnetic interference (EMI) from the EMS device caused the delivery of an inappropriate ICD discharge. In a second patient with an ICD with endocardial true bipolar sensing, there was no evidence of EMI during the EMS therapy despite all of our attempts to reproduce it. The sensing circuits in the two different ICDs are compared.
No takes yet. Share an insight, caveat, or question.
Glotzer et al. (1998) studied this question.