Why the study?
Does oscillometric blood pressure measurement differ from auscultatory blood pressure measurement in schoolchildren?
Does oscillometric blood pressure measurement differ from auscultatory blood pressure measurement in schoolchildren?
Answers are grounded in this paper and cite their sources.
Oscillometric blood pressure devices yield higher readings than auscultatory methods in children, indicating the two methods cannot be used interchangeably for diagnosing hypertension.
OBJECTIVE: To study the differences in blood pressure readings between the auscultatory and oscillometric (Dinamap model 8100; Critikon, Tampa, Fla) methods. DESIGN: Survey of 2 blood pressure instruments. SETTING: Public schools. PARTICIPANTS: Seven thousand two hundred eight schoolchildren aged 5 through 17 years. MAIN OUTCOME MEASURE: Blood pressure levels. RESULTS: For all children combined, Dinamap systolic pressure readings were 10 mm Hg higher (95% confidence interval, -4 to 24 mm Hg) than the auscultatory systolic pressure readings. Dinamap diastolic pressure readings were 5 mm Hg higher (95% confidence interval, -14 to 23 mm Hg) than the auscultatory Korotkoff phase V diastolic pressure readings. CONCLUSION: These findings preclude the interchange of readings by the 2 methods. Caution must be exercised in the diagnosis of hypertension when an automated device is used.
No takes yet. Share an insight, caveat, or question.
Park et al. (2001) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: