Key result
A modified radioimmunoassay for digoxin using an initial dichlormethane extraction achieved a precision of 5-7% and extracted more than 93% of digoxin.
A modified radioimmunoassay provides a precise method for measuring digoxin levels in various biological samples, suitable for pharmacokinetic studies.
Answers are grounded in this paper and cite their sources.
May facilitate pharmacokinetic digoxin studies; leaves open clinical validation before routine use.
A modified radioimmunoassay for digoxin, suitable for serum, urine and homogenized myocardial tissue, is presented. It includes an initial extraction by dichlormethane after addition of protein (gelatine, 4 g/l). A constant fraction and more than 93 per cent of the digoxin is extracted in this way. The method is suitable for pharmacokinetic studies, since the basis for standardization is aqueous solutions of digoxin, and the interference of proteins in the samples are eliminated. By this means the interference of other digitalis glycosides is reduced, and quenching correction under the final determination of the 3H activity is unnecessary. The precision, including the initial extraction, is 5–7 per cent.
No takes yet. Share an insight, caveat, or question.
Axel Brock (1974) studied this question. Modified radioimmunoassay for digoxin was evaluated on Precision of the assay. A modified radioimmunoassay for digoxin using an initial dichlormethane extraction achieved a precision of 5-7% and extracted more than 93% of digoxin.
Synapse has enriched 10 related papers. Consider them for comparative context: