Key result
Progressive enlargement of the Silastic ball component of a Starr-Edwards aortic prosthesis caused severe hemolytic anemia, which disappeared after the prosthesis was replaced with a homograft valve.
Case Report (n=1)
Answers are grounded in this paper and cite their sources.
Highlights a rare complication of Starr-Edwards aortic valve prostheses where lipid deposition causes ball enlargement, obstruction, and hemolytic anemia, which is curable by valve replacement.
The patient described developed hemolytic anemia following insertion of a Starr-Edwards aortic valve prosthesis. No diastolic murmur was heard and no insufficiency was detected. The hemolytic anemia was progressively more severe and uncompensated despite various medical measures. The anemia was characterized by fragmented erythrocytes in the peripheral blood, reticulocytosis, elevation of plasma heme pigments augmented by exercise, increased fecal urobilinogen, and iron loss in the urine. Studies of chromium-51-tagged erythrocytes indicated an extracorpuscular mechanism of hemolysis. At reoperation the ball component of the prosthesis was found to be enlarged and obstructing blood flow. Chemical analysis of the ball showed significant cholesterol and lipid deposits. The hemolysis disappeared after the prosthesis was replaced with a homograft valve. Ferrokinetic studies showed rapid plasma clearance and incorporation into circulating erythrocytes. Triglycerides were elevated during the period of hemolysis. Red cell membrane lipids were normal.
No takes yet. Share an insight, caveat, or question.
GARCIA et al. (1968) conducted a case report in Hemolytic anemia (n=1). Starr-Edwards aortic valve prosthesis was evaluated on Hemolysis resolution. Progressive enlargement of the Silastic ball component of a Starr-Edwards aortic prosthesis caused severe hemolytic anemia, which disappeared after the prosthesis was replaced with a homograft valve.
Synapse has enriched 10 related papers. Consider them for comparative context: