Answers are grounded in this paper and cite their sources.
A torque meter is described which employs the principles used in work with resistance strain gauges to convert small torques developed in isolated specimens directly into electrical signals. The overall sensitivity and the maximum torque that can be measured with this meter can both be altered appreciably by suitable modifications in design, but for the instrument described the overall sensitivity is about 3 dyn cm per cm deflection and the maximum torque is approximately 4 × 10 3 dyn cm. The meter has been used to measure the anisotropy torque developed in a (111) single-crystal disk of nickel and hence the second magnetocrystalline anistropy constant K 2 . At 23° C the effective values of K 2 appropriate to 7300 Oe and infinite field are respectively 2.38 × 10 4 and 2.53 × 10 4 erg cm -3 . These results are in excellent agreement with the direct determination of K 2 for nickel by Sato and Chandrasekhar.
No takes yet. Share an insight, caveat, or question.
Birss et al. (1963) studied this question.