Key result
Serum digoxin concentration measured by EIA was higher than by FPIA.
Why the study?
Does dry chemistry EIA on the Vitros 950 yield different serum digoxin concentrations compared to FPIA on the TDxFLx in patient samples?
Observational
Does dry chemistry EIA on the Vitros 950 yield different serum digoxin concentrations compared to FPIA on the TDxFLx in patient samples?
Significant method-dependent differences exist between EIA and FPIA for measuring serum digoxin, necessitating harmonization if both systems are used in the same clinical setting.
Answers are grounded in this paper and cite their sources.
Method bias may alter digoxin interpretation across labs; leaves open outcome impact and harmonization needs.
The aim of the study was to compare the results of digoxin assays performed using fluorescence polarization immunoassay (FPIA) on the TDxFLx and a dry chemistry enzyme immunoassay (EIA) on the Vitros 950. Within-run CV amounted to 8.52-10.49% for FPIA and 2.47-5.39% for EIA. Between-run CV amounted to 6.41-8.97% for FPIA and 3.40-5.04% for EIA. Analytical bias ranged from 2.57-4.0% for FPIA and from 9.86-11.9% for EIA. In comparative studies the correlation coefficient was 0.878; Deming regression analysis yielded a slope of 1.057 (95% CI: 0.573 to 1.541) and intercept of 0.078 (95% CI: -0.391 to 0.547), and the Passing-Bablok agreement test yielded a slope of 1.111 (95% CI: 0.988 to 1.212) and intercept of 0.094 (95% CI: -0.018 to 0.182). The mean digoxin concentration in patients' sera measured by EIA was significantly higher than that measured by FPIA (1.347 vs. 1.196 ng/ml, p<0.02). The mean absolute difference between results amounted to 0.146 ng/ml (95% CI: 0.0261 to 0.266). In comparison to EIA, FPIA yielded a higher number of subtherapeutic concentrations <0.5 ng/ml (29.7% vs. 21.8%) and a lower number of digoxin concentrations >1.2 ng/ml (25.7% vs. 35.6%). These discrepancies occurred in approximately 10% of samples. The obtained results showed different analytical performance and method-dependent differences in the distribution of results. This indicates the necessity to harmonize digoxin immunoassays if two different analytical systems are used in the same clinical setting.
No takes yet. Share an insight, caveat, or question.
Bogdan Solnica (2004) conducted an observational in Serum digoxin concentration monitoring. Dry chemistry enzyme immunoassay (EIA) vs. Fluorescence polarization immunoassay (FPIA) was evaluated on Mean digoxin concentration (MD 0.146, 95% CI 0.0261 to 0.266, p=<0.02). Serum digoxin concentration measured by EIA was higher than by FPIA.