Key result
A bio-inspired fiber-reinforced aortic valve scaffold demonstrated adequate acute mechanical and hemodynamic properties, with a transvalvular pressure gradient of 3.02 ± 0.26 mmHg and EOA of 2.12 cm2.
Why the study?
The application of tissue-engineered heart valves in the high-pressure circulatory system remains challenging.
A novel manufacturing approach for fiber-reinforced biomimetic scaffolds for aortic heart valves demonstrated adequate acute mechanical and hemodynamic properties suitable for in situ valve replacement.
Biohybrid valve scaffolds merit preclinical testing; extends textile reinforcement concepts but leaves clinical durability open.
The application of tissue-engineered heart valves in the high-pressure circulatory system is still challenging. One possible solution is the development of biohybrid scaffolds with textile reinforcement to achieve improved mechanical properties. In this article, we present a manufacturing process of bio-inspired fiber reinforcement for an aortic valve scaffold. The reinforcement structure consists of polyvinylidene difluoride monofilament fibers that are biomimetically arranged by a novel winding process. The fibers were embedded and fixated into electrospun polycarbonate urethane on a cylindrical collector. The scaffold was characterized by biaxial tensile strength, bending stiffness, burst pressure and hemodynamically in a mock circulation system. The produced fiber-reinforced scaffold showed adequate acute mechanical and hemodynamic properties. The transvalvular pressure gradient was 3.02 ± 0.26 mmHg with an effective orifice area of 2.12 ± 0.22 cm2. The valves sustained aortic conditions, fulfilling the ISO-5840 standards. The fiber-reinforced scaffold failed in a circumferential direction at a stress of 461.64 ± 58.87 N/m and a strain of 49.43 ± 7.53%. These values were above the levels of tested native heart valve tissue. Overall, we demonstrated a novel manufacturing approach to develop a fiber-reinforced biomimetic scaffold for aortic heart valve tissue engineering. The characterization showed that this approach is promising for an in situ valve replacement.
No takes yet. Share an insight, caveat, or question.
Boehm et al. (2023) studied Aortic valve replacement. Bio-inspired fiber reinforcement for an aortic valve scaffold vs. Native heart valve tissue was evaluated on Mechanical and hemodynamic properties. A bio-inspired fiber-reinforced aortic valve scaffold demonstrated adequate acute mechanical and hemodynamic properties, with a transvalvular pressure gradient of 3.02 ± 0.26 mmHg and EOA of 2.12 cm2.