This note describes a specimen airlock chamber designed to replace, without modification, the existing airlock of the Siemens Elmiskop I. In addition to the standard specimen holders the airlock will load and unload a goniometer hot stage having a maximum temperature of 720°C and a tilting angle of ±25°. A decontamination device is also included together with an electromagnetic beam tilting system. The decontamination device decreases by more than 50 the specimen contamination rate normally found in the Siemens Elmiskop I, and the electromagnetic coils tilt the beam 5° in any direction for the application of dark field electron microscopy.