Key result
MediGuide sensor-based navigation during CRT implant procedures reduced total ionizing radiation exposure by 66% (median 769 vs 2,608 μGray·m2; P<0.001) compared to standard procedures.
Why the study?
Does a sensor-based electromagnetic tracking system reduce ionizing radiation exposure during CRT implant procedures?
Cohort (n=130)
Does a sensor-based electromagnetic tracking system reduce ionizing radiation exposure during CRT implant procedures?
Sensor-based electromagnetic tracking systems significantly reduce ionizing radiation exposure during CRT implant procedures, primarily by facilitating coronary sinus cannulation.
Answers are grounded in this paper and cite their sources.
May support radiation minimization in CRT implants; leaves open need for RCTs before practice change.
BACKGROUND: Cardiac resynchronization therapy (CRT) implant procedures are often complex and prolonged, resulting in significant ionizing radiation (IR) exposure to the patient and operator. We report our early experience working with a novel sensor-based electromagnetic tracking system (MediGuide™, MDG, St. Jude Medical Inc., St. Paul, MN, USA), in terms of procedural IR exposure reduction. METHODS AND RESULTS: Information regarding patient demographics, procedural details, procedural duration, and IR exposure were prospectively collected on 130 consecutive CRT procedures performed between January 2013 and January 2014. Sixty procedures were performed with MDG guidance, and 70 were performed without MDG guidance. Despite a nonsignificant trend toward shorter procedure duration with the use of MDG (120 minutes vs 138 minutes with non-MDG, P = 0.088), a 66% reduction in total IR exposure (median 769 μGray · m(2) vs 2,608 μGray · m(2), P < 0.001) was found. This reduction was primarily driven by a >90% reduction in IR dose required to cannulate the coronary sinus (median 80 μGray · m(2) vs 922 μGray · m(2), P < 0.001), and to a lesser extent from a reduction in IR dose required for LV lead placement (median 330 μGray·m(2) vs 737 μGray · m(2), P = 0.059). In addition, a significant learning curve effect was observed with a significantly shorter procedural duration for the last 15 cases compared to the first 15 cases (median 98 minutes vs 175 minutes, P < 0.001). CONCLUSION: The nonfluoroscopic MDG positioning system is associated with a dramatic reduction in exposure to IR during CRT implant procedures, with a 90% decrease in the IR dose required to cannulate the coronary sinus. A steep learning curve was quantified.
No takes yet. Share an insight, caveat, or question.
Thibault et al. (2014) conducted a cohort in Cardiac resynchronization therapy (CRT) implant (n=130). MediGuide sensor-based electromagnetic tracking system vs. Without MediGuide guidance was evaluated on Total ionizing radiation (IR) exposure (66% reduction, p=<0.001). MediGuide sensor-based navigation during CRT implant procedures reduced total ionizing radiation exposure by 66% (median 769 vs 2,608 μGray·m2; P<0.001) compared to standard procedures.
Synapse has enriched 10 related papers. Consider them for comparative context: