Key result
Use of the MediGuide Technology sensor-based tracking system for CRT implantation achieved 100% successful left ventricular lead placement with a median total fluoroscopy time of 5.2 minutes.
Why the study?
Does a novel sensor-based electromagnetic tracking system allow for safe and feasible CRT device implantation in patients with an established indication for CRT?
Does a novel sensor-based electromagnetic tracking system allow for safe and feasible CRT device implantation in patients with an established indication for CRT?
The use of a novel sensor-based electromagnetic tracking system for CRT implantation is feasible and safe, demonstrating potential for reduced fluoroscopy time.
Answers are grounded in this paper and cite their sources.
Supports feasibility of sensor-based CRT implantation with low fluoroscopy; hypothesis-generating and requires validation in prospective trials.
BACKGROUND: Cardiac resynchronization therapy (CRT) device implantation can be challenging, time consuming, and fluoroscopy intense. To facilitate left ventricular lead placement, a novel sensor-based electromagnetic tracking system (MediGuide Technology [MGT], St. Jude Medical) has been developed. We report the results of the First Human Use study evaluating the feasibility, safety, and performance of a novel CRT implantation approach using electromagnetic trackable operation equipment. METHODS AND RESULTS: Fifteen consecutive patients (66±8 years, 53% male) with an established indication for CRT were implanted using the new tracking technology. Demographics, anatomical information, detailed fluoroscopy need, procedure time, and adverse events were collected. Patients were followed up for 4 weeks after implantation. The CRT system was successfully implanted with a lateral or posterolateral left ventricular lead position in all patients. The total procedure time was 116±43 minutes, the median total fluoroscopy time (skin to skin) was 5.2 (Q1-Q3, 3.0-8.4) minutes, and the median fluoroscopy time for left ventricular lead deployment (coronary sinus [CS] cannulation to withdrawal of CS sheath) measured 2.6 (Q1-Q3, 1.6-5.6) minutes. There were no severe complications that required an acute intervention or reoperation during the perioperative and postoperative periods. CONCLUSIONS: Use of the MGT tracking technology allows for safe and successful CRT implantation with the potential for reduced fluoroscopy time. Future randomized studies are needed to validate these data. CLINICAL TRIAL REGISTRATION: URL http://www.clinicaltrials.gov. Unique identifier: NCT01519739.
No takes yet. Share an insight, caveat, or question.
Richter et al. (2013) studied established indication for cardiac resynchronization therapy (n=15). MediGuide Technology sensor-based electromagnetic tracking system was evaluated on successful implantation with a lateral or posterolateral left ventricular lead position. Use of the MediGuide Technology sensor-based tracking system for CRT implantation achieved 100% successful left ventricular lead placement with a median total fluoroscopy time of 5.2 minutes.
Synapse has enriched 10 related papers. Consider them for comparative context: