Key result
The OMRON HEM-907XL device was accurate for diastolic BP but less accurate for systolic BP compared to aneroid auscultation in CKD patients.
Why the study?
Does the OMRON HEM-907XL oscillometric blood pressure monitor accurately measure blood pressure compared to auscultation in nondialytic patients with chronic kidney disease?
Observational (n=87)
Blinded observers
Does the OMRON HEM-907XL oscillometric blood pressure monitor accurately measure blood pressure compared to auscultation in nondialytic patients with chronic kidney disease?
The OMRON HEM-907XL monitor is accurate for measuring diastolic BP but performs less well for systolic BP in nondialytic CKD patients.
Answers are grounded in this paper and cite their sources.
Supports diastolic BP use in nondialytic CKD; leaves open systolic accuracy concerns for prospective validation.
The OMRON HEM-907XL is a commercial oscillometric blood pressure (BP) monitor that was used in the Systolic Blood Pressure Intervention Trial (SPRINT), in which 28% of participants had chronic kidney disease (CKD). This study examined the accuracy of the monitor in nondialytic patients with CKD. Eighty-seven patients met inclusion criteria. The authors used a modified Association for the Advancement of Medical Instrumentation (AAMI) protocol, with one observer recording measurements from the monitor and two blinded physicians obtaining simultaneous aneroid values by auscultation. Using AAMI method 1, there was a 2.5±9.5 mm Hg difference in OMRON and aneroid systolic BP, and a -1.6±6.5 mm Hg difference in diastolic BP. Using AAMI method 2, there was a 5.1±7.4 mm Hg difference in systolic BP and a -0.2±5.4 mm Hg difference in diastolic BP. In patients with CKD, the OMRON HEM-907XL appears to be accurate for measuring diastolic BP, but did not perform as well for systolic BP.
No takes yet. Share an insight, caveat, or question.
Cohen et al. (2017) conducted an observational in Nondialytic chronic kidney disease (n=87). OMRON HEM-907XL vs. Aneroid auscultation was evaluated on Difference in systolic and diastolic blood pressure (MD 2.5). The OMRON HEM-907XL device was accurate for diastolic BP but less accurate for systolic BP compared to aneroid auscultation in CKD patients.