The study presents a new anatomically adapted design for an electromechanical total artificial heart with angled pump chambers and a novel mechanical gear.
Answers are grounded in this paper and cite their sources.
Oblique chamber orientation may improve anatomical fit in total artificial hearts; leaves open in vivo hemodynamic and durability validation.
The intrathoracic anatomical situation after explantation of the natural heart defines the maximum available space for the design of the housing as well as of the inlet- and outlet connectors of a fully implantable electromechanical artificial heart. Based on computer-assisted anatomical studies, a total artificial heart housing is designed which facilitates an oblique orientation of the pumping chambers for a better fluidmechanical and anatomical arrangement of the in- and outlet connectors. The pumping chamber geometry is based on modifications of an existing cardiac assist-system. Subsequently a mechanical gear which conforms to this anatomically adapted housing is developed.
No takes yet. Share an insight, caveat, or question.
Kaufmann et al. (1992) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: