Key result
A pump circuit using a reversible rotary gear pump and 4 artificial heart valves achieved a measured flow rate of 0.8-0.75 L/min at a heart rate of 60-110 bpm.
A novel reversible rotary gear pump circuit with four artificial valves demonstrated the ability to generate pulsatile flow alternately into systemic and pulmonary circuits in a benchtop model.
Answers are grounded in this paper and cite their sources.
May enable compact pulsatile artificial hearts; extends benchtop evidence but leaves open in vivo validation.
A pump circuit was assembled and examined for use as an implantable artificial heart. The circuit consisted of a gear pump and 4 artificial heart valves. Mitral and pulmonary arterial valves were placed at the inflow port of the pump, and aortic and tricuspid valves were placed at the outflow port. The mitral and the tricuspid valves were connected to each reservoir at 10 mm Hg, and the aortic and the pulmonary arterial valves were connected to the head tanks at 80 and 20 mm Hg, respectively. The pump discharged pulsatile flows into both systemic and pulmonary arteries alternately by switching the direction of rotation periodically. Because the rated discharge was 1.7 L/min for the gear pump used, the measured flow rate was 0.8-0.75 L/min at a heart rate of 60-110 bpm.
No takes yet. Share an insight, caveat, or question.
Fukunaga et al. (1995) studied this question. Pulsatile total artificial heart using a reversible rotary pump was evaluated on Measured flow rate. A pump circuit using a reversible rotary gear pump and 4 artificial heart valves achieved a measured flow rate of 0.8-0.75 L/min at a heart rate of 60-110 bpm.