This is the authors' abstract. We don't add key points for this paper.
A reverse phase, isocratic HPLC method has been developed for the quantitation of quinine and its major metabolite, 3-hydroxyquinine in human plasma, urine and hepatic microsomal samples. The method involves simple protein precipitation for sample treatment and ion-paired chromatography. The chromatographic separation is accomplished with a mobile phase comprising acetonitrile-aqueous phosphate buffer (40:60, v/v) containing 10 mM sodium dodecyl sulphate and 0.1 mM tetrabutylammonium bromide and adjusted to pH 2.1. The mobile phase is pumped at a flow rate of 0.5 mL/min. A microbore column is used (2 mm I.D. × 100 mm) packed with a C18 reverse phase material (5 μm ODS Hypersil). Biological samples (100–500 μL) were precipitated with two volumes of cold methanol, vortexed and then centrifuged at 1500 g for 10 min. The supernatant (30 μL) was injected into the HPLC column. The chromatograms were monitored using a fluorescence detector setting with excitation and emission wavelenths of 350 and 450 nm, respectively. Under these conditions, the lower limit of detection was 0.1 μM (0.034 μg/mL) for the major metabolite 3-hydroxyquinine, and 0.5 μM (0.16 μg/mL) for quinine. The inter- and intra-assay coefficients of variation were found to be less than 7%. The assay procedure is applicable for studying the pharmacokinetics and metabolism of quinine.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Wanwimolruk et al. (1996) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: