Key result
A liquid chromatography method was developed that accurately measures quinidine in serum and urine to levels as low as /L without interference from metabolites or common drugs.
A liquid chromatography method was developed to accurately measure quinidine in serum and urine down to 0.5 mg/L without interference.
Answers are grounded in this paper and cite their sources.
May facilitate quinidine TDM; leaves open clinical adoption or outcome impact pending validation.
A liquid chromatography method for the determination quinidine in serum and urine has been developed. This method uses isocratic conditions, ambient temperature, a conventional fixed wavelength, 254-nm detector, and is free of potential interference from quinidine metabolites. Sample pretreatment involves extraction of quinidine along with quinine, an internal standard, into an organic phase and reextraction into an aqueous acidic phase. In this manner, interference due to commonly used drugs are eliminated. The method is capable of accurately measuring quinidine to levels as low as 0.5 mg/L.
No takes yet. Share an insight, caveat, or question.
C. P. Patel (1982) studied this question. Liquid chromatography method was evaluated on Accurate measurement of quinidine. A liquid chromatography method was developed that accurately measures quinidine in serum and urine to levels as low as /L without interference from metabolites or common drugs.