The described liquid chromatography method for determining plasma quinidine concentrations is highly reproducible and correlates well with established fluorometric procedures.
Answers are grounded in this paper and cite their sources.
May support LC for quinidine monitoring; leaves open prospective outcome validation before routine adoption.
We describe a routine method for determining plasma concentrations of quinidine by liquid chromatography. The procedure requires 1.0 ml of plasma (or serum) and involves internal standard addition, extraction with ether, and separation on a column of microparticulate silica. Day-to-day CV (15 days) was less than 5% and no deterioration in column performance has been observed during 12 months. Comparison with a fluorometric procedure gave a correlation coefficient of 0.995.
No takes yet. Share an insight, caveat, or question.
Peat et al. (1978) studied this question.