Answers are grounded in this paper and cite their sources.
A motor rotation in an electromechanically-driven pulsatile total artificial heart (TAH) may influence hemolysis. This study is designed to evaluate motor rotational conditions of the TAH and choose a suitable condition to obtain the least hemolytic characteristics. The TAH was driven in two motor rotational conditions: a constant motor rotational speed (rpm) mode (mode A) and a gradually increasing rpm mode (mode B). In these two modes, a maximum dP/dt value and a degree of hemolysis were measured and compared. The TAH was connected to an in vitro testing loop. In each mode, the TAH was driven with a fixed pumping rate of 100 bpm. A preload and an afterload were held at 15 and 100 mm Hg, respectively. The outflow of the TAH was maintained at 4.0 L/min. The maximum dP/dt in mode A and mode B was 5914 +/- 405 mm Hg/s and 2953 +/- 191 mm Hg/s, respectively. The NIH value obtained from mode A and mode B was 0.063 +/- 0.005 g/100 L and 0.026 +/- 0.003 g/100 L, respectively. The results demonstrated that the TAH driven in a gradually increasing rpm mode reduces both a maximum dP/dt value and a degree of hemolysis. The gradually increasing rpm mode is a suitable driving condition to obtain the least hemolytic characteristics.
No takes yet. Share an insight, caveat, or question.
Ohashi et al. (2003) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: