Key result
Hemolysis in a mock circulatory loop with normal aortic blood pressure was significantly higher than with lowered aortic pressure, and none of the tested chemical additives significantly reduced hemolysis.
In vitro testing using a novel mock circulatory loop showed that EVAD-induced hemolysis is pressure-dependent and not mitigated by common drag-reducing agents.
Answers are grounded in this paper and cite their sources.
May warrant pressure-lowering strategies in EVADs; leaves open additive efficacy pending in vivo validation.
A new mock circulatory loop was developed for hemolysis studies associated with the Penn State electric ventricular assist device (EVAD). This flow loop has several advantages over previously designed loops. It is small enough to accommodate experiments in which only single units of blood are available, it is made out of biocompatible materials, it incorporates good geometry, and it provides normal physiological pressures and flows to both the aortic outlet and the venous inlet of the pumping device. Experiments with reduced aortic pressure but normal cardiac output showed that hemolysis in a loop with normal aortic blood pressure was significantly higher than that in a loop with lowered aortic pressure, thereby illustrating the importance of maintaining loop pressures as close as possible to those found in vivo. This data also imply that blood traveling through the left ventricle in an artificial heart may be subject to higher hemolysis rates than that traversing the right ventricle. Another set of experiments to determine the effects of 4 hemolysis or drag-reducing agents (Pluronic F-68, Dextran-40, Polyox WSR-301, and Praestol 2273TR) on blood trauma due to the EVAD and associated valves was performed. Results indicated that none of the additives significantly reduced hemolysis under the conditions found in the mock loop. Finally, a compilation of data gathered in these experiments showed that the index of hemolysis (IH) is dependent on hematocrit (HCT), which suggests that another parameter, IH/HCT, may be more suited to the quantification of hemolysis.
No takes yet. Share an insight, caveat, or question.
Garrison et al. (1994) studied Hemolysis associated with ventricular assist devices. Chemical additives and varying aortic pressures vs. Normal aortic pressure / no additives was evaluated on Hemolysis (index of hemolysis). Hemolysis in a mock circulatory loop with normal aortic blood pressure was significantly higher than with lowered aortic pressure, and none of the tested chemical additives significantly reduced hemolysis.
Synapse has enriched 10 related papers. Consider them for comparative context: