Key result
A small-diameter interatrial shunt in a totally implantable artificial heart achieved excellent right-left ventricular output balance, maintaining left atrial pressure 5 mmHg above right atrial pressure.
An interatrial shunt in a totally implantable artificial heart effectively balances right and left atrial pressures, eliminating the need for bulky volume-displacement chambers.
Answers are grounded in this paper and cite their sources.
May enable simpler totally implantable artificial hearts; leaves open human translation from this animal model.
Pneumatically powered artificial hearts readily accommodated the higher net stroke volumes by the right ventricle than from the left ventricle. We published that this discrepancy was approximately 8% of the left ventricular cardiac output. A variety of methods have been used to achieve balance between the right and left atrial pressures. Relatively large volume-displacement chambers (VDC) present potential problems, but do provide balance. The VDC in volumetrically coupled right-left stroke volumes was eliminated by using a small-diameter interatrial shunt (IAS). Preliminary studies demonstrated excellent balance in contracted and expanded blood volume (preload) and by hypotension and hypertension created with vasoactive drugs (afterload). At a mean aortic pressure of 120 mmHg, heart rate of 120 BPM, cardiac output of 8 L/minute and right atrial pressure of 13 mmHg, the peak IAS flow was 3.2 ml/beat in a right to left direction and 8.0 ml/beat in a left to right direction. The net left to right flow was 4.8 ml/beat. Over a wide range of preload (2 to 20 mmHg) and afterload (45 to 180 mmHg), the left atrial pressure was routinely 5 mm Hg more than the right atrial pressure. Elimination of the VDC reduces the number of components, volume, and weight of the totally implantable artificial heart. The IAS offers a simple solution to a very complex problem and provides a device that is simpler to implant and is possible to explant.
No takes yet. Share an insight, caveat, or question.
Olsen et al. (1991) studied Right-left ventricular output balance in artificial hearts. Small-diameter interatrial shunt (IAS) vs. Volume-displacement chambers (VDC) was evaluated on Right-left ventricular output balance (net flow and atrial pressures). A small-diameter interatrial shunt in a totally implantable artificial heart achieved excellent right-left ventricular output balance, maintaining left atrial pressure 5 mmHg above right atrial pressure.
Synapse has enriched 10 related papers. Consider them for comparative context: