Key result
A microfluidic chip-based hydrodynamic focusing approach successfully quantified platelet surface coverage and aggregate size distributions using ~50 μL of whole blood in a 2-minute assay.
A novel microfluidic approach enables rapid, low-volume assessment of platelet adhesion under arterial shear conditions, with potential applications for point-of-care diagnostics.
Answers are grounded in this paper and cite their sources.
May advance preclinical platelet assays with minimal volumes; leaves open human validation for point-of-care use.
We report a microfluidic chip-based hydrodynamic focusing approach that minimizes sample volume for the analysis of cell-surface interactions under controlled fluid-shear conditions. Assays of statistically meaningful numbers of translocating platelets interacting with immobilized von Willebrand factor at arterial shear rates (∼1500 s(-1)) are demonstrated. By controlling spatial disposition and relative flow rates of two contacting fluid streams, e.g., sample (blood) and aqueous buffer, on-chip hydrodynamic focusing guides the cell-containing stream across the protein surface as a thin fluid layer, consuming ∼50 μL of undiluted whole blood for a 2-min platelet assay. Control of wall shear stress is independent of sample consumption for a given flow time. The device design implements a mass-manufacturable fabrication approach. Fluorescent labeling of cells enables readout using standard microscopy tools. Customized image-analysis software rapidly quantifies cellular surface coverage and aggregate size distributions as a function of time during blood-flow analyses, facilitating assessment of drug treatment efficacy or diagnosis of disease state.
No takes yet. Share an insight, caveat, or question.
Kent et al. (2010) studied this question. Microfluidic chip-based hydrodynamic focusing approach was evaluated on Cellular surface coverage and aggregate size distributions. A microfluidic chip-based hydrodynamic focusing approach successfully quantified platelet surface coverage and aggregate size distributions using ~50 μL of whole blood in a 2-minute assay.
Synapse has enriched 10 related papers. Consider them for comparative context: