Key result
CORS test demonstrates strong reproducibility for cardiac output with a ~7% coefficient of variation.
Why the study?
Does the novel non-invasive Cardiac Output Response to Stress (CORS) test provide reproducible haemodynamic and metabolic measurements at rest and during step-exercise in older primary care patients?
Observational (n=32)
No
Does the novel non-invasive Cardiac Output Response to Stress (CORS) test provide reproducible haemodynamic and metabolic measurements at rest and during step-exercise in older primary care patients?
Answers are grounded in this paper and cite their sources.
The novel non-invasive bioreactance-based Cardiac Output Response to Stress (CORS) test demonstrates acceptable test-retest reproducibility (CV 7.1%, r=0.92), supporting further evaluation of its diagnostic utility as a rule-in test for heart failure in primary care.
AIMS: Primary care physicians lack access to an objective cardiac function test. This study for the first time describes a novel cardiac output response to stress (CORS) test developed to improve diagnosis and monitoring of heart failure in primary care and investigates its reproducibility. METHODS AND RESULTS: Prospective observational study recruited 32 consecutive primary care patients (age, 63 ± 9 years; female, n = 18). Cardiac output was measured continuously using the bioreactance method in supine and standing positions and during two 3 min stages of a step-exercise protocol (10 and 15 steps per minute) using a 15 cm height bench. The CORS test was performed on two occasions, i.e. Test 1 and Test 2. There was no significant difference between repeated measures of cardiac output and stroke volume at supine standing and Stage 1 and Stage 2 step exercises (all P > 0.3). There was a significant positive relationship between Test 1 and Test 2 cardiac outputs (r = 0.92, P = 0.01 with coefficient of variation of 7.1%). The mean difference in cardiac output (with upper and lower limits of agreement) between Test 1 and Test 2 was 0.1 (-1.9 to 2.1) L/min, combining supine, standing, and step-exercise data. CONCLUSIONS: The CORS, as a novel test for objective evaluation of cardiac function, demonstrates acceptable reproducibility and can potentially be implemented in primary care.
No takes yet. Share an insight, caveat, or question.
Charman et al. (2018) conducted an observational in Primary care patients without chronic, cardiovascular, pulmonary, or metabolic diseases (n=32). Cardiac Output Response to Stress (CORS) test using non-invasive bioreactance technology vs. Repeated CORS test (Test 1 vs Test 2 separated by at least 1 hour of sitting rest) was evaluated on Reproducibility of cardiac output measurements between Test 1 and Test 2 across combined rest, standing, and step-exercise stages (r = 0.92, CV 7.1%, MD 0.1 L/min (limits of agreement -1.9 to 2.1), p=0.01). The Cardiac Output Response to Stress (CORS) test demonstrated acceptable reproducibility between repeated tests with a strong positive correlation in cardiac output.
Synapse has enriched 10 related papers. Consider them for comparative context: