Key result
Side-by-side placement of two polyurethane leads through a restrictive costoclavicular space resulted in simultaneous insulation deterioration at 11 months and complete lead failure at 16 months.
Case Report (n=1)
Insertion of polyurethane leads through a tight costoclavicular region should be avoided to prevent delayed lead failure from mechanical stress.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating for insulation failure with side-by-side polyurethane leads in tight costoclavicular space; prospective data needed before practice change.
During implantation of a bipolar dual chamber pacing system, two polyurethane leads were placed side-by-side through a single left subclavian vein puncture. The costoclavicular space was restrictive at the insertion site, and binding of the leads in this region made positioning difficult. Eleven months later, although pacemaker function remained normal, deterioration of both leads' insulation was noted on routine chest roentgenogram. Sixteen months after implantation, both leads failed. Insertion of polyurethane leads through a tight costoclavicular region should be avoided as subsequent mechanical stressing of the insulation may result in delayed lead failure.
No takes yet. Share an insight, caveat, or question.
F. Earl Fyke (1988) conducted a case report in Pacemaker lead implantation (n=1). Side-by-side subclavian placement of two polyurethane leads was evaluated on Lead failure and insulation deterioration. Side-by-side placement of two polyurethane leads through a restrictive costoclavicular space resulted in simultaneous insulation deterioration at 11 months and complete lead failure at 16 months.