Bipolar polyurethane leads (Medtronic 6972) implanted in the right ventricle resulted in insulation failure in 6.6% of cases at 15 to 18 months post-implant.
Observational (n=45)
No
Answers are grounded in this paper and cite their sources.
Bipolar polyurethane pacing leads (Medtronic 6972) demonstrated a 6.6% rate of insulation failure at 15 to 18 months post-implant, characterized by low impedance and inappropriate pacemaker inhibition without visible cracking.
Between October, 1981, and April, 1983, 200 pacemakers were inserted at Rhode Island Hospital. Forty-five bipolar polyurethane leads (Medtronic 6972) were positioned in the right ventricle. Three (6.6%) of these leads presented 15 to 18 months post-implant with insulation failure. The manifestation was inappropriate inhibition of the pacemaker. No obvious cracking was visible on X-ray or during surgery. All the leads had low impedance, indicative of insulation failure.
No takes yet. Share an insight, caveat, or question.
Raymond et al. (1984) conducted an observational in Pacemaker implantation (n=45). Bipolar polyurethane leads (Medtronic 6972) was evaluated on Insulation failure. Bipolar polyurethane leads (Medtronic 6972) implanted in the right ventricle resulted in insulation failure in 6.6% of cases at 15 to 18 months post-implant.