Key result
ThinLine coradial bipolar leads demonstrated lower capture thresholds than coaxial controls, comparable sensing, and acceptable impedances over 17,530 device months.
Why the study?
Do ThinLine coradial bipolar leads improve pacing thresholds and safety compared to conventional coaxial leads in patients requiring cardiac pacing?
RCT (n=1,068)
Randomized (intensive phase)
Yes
Do ThinLine coradial bipolar leads improve pacing thresholds and safety compared to conventional coaxial leads in patients requiring cardiac pacing?
ThinLine coradial bipolar leads demonstrate safety, efficacy, and favorable handling characteristics with lower capture thresholds compared to conventional coaxial leads.
Answers are grounded in this paper and cite their sources.
May support ThinLine leads for lower thresholds in paced patients; extends observational comparisons but leaves open randomized confirmation.
Bipolar, transvenous, passive-fixation leads (ThinLine, Sulzer Intermedics Inc., Angleton, TX, USA) incorporating coradial individually coated conductor coils, and a redundant external 55D polyurethane insulation sheath were developed. The diameter of the new leads (< 5 Fr) is in the range of available unipolar leads and is considerably smaller than conventional bipolar pacing leads. From January 9, 1994 to November 12, 1996, 1,536 model 432-04 (523 atrial) and 430-10 (1,013 ventricular) leads were implanted in 1,068 patients at 50 US and 5 Canadian centers to evaluate their safety and efficacy. The study included a general phase, with follow-ups at 1, 3, 6, 12, 18, and 24 months; and a randomized intensive phase with collection of more data and additional 2- and 6-week follow-ups. Capture and sensing thresholds, lead impedance, and handling characteristics were evaluated. Clinical events were monitored and performance was compared to that of two commercially available conventional (coaxial) leads: Sulzer Intermedics models 432-03 (atrial) and 430-07 (ventricular). During a total of 17,530 device months, there were two lead failures, no lead related deaths, 32 explants, 37 complications, and no unexpected adverse device effects. Capture thresholds were lower than those for coaxial controls with identical electrodes, sensing was comparable, lead impedances were within clinically acceptable ranges, and investigators found overall handling characteristics good to excellent. ThinLine coradial bipolar leads are safe and effective for cardiac pacing and sensing.
No takes yet. Share an insight, caveat, or question.
Belott et al. (1998) conducted an RCT in Cardiac pacing (n=1,068). ThinLine coradial bipolar leads vs. Conventional coaxial leads was evaluated on Safety and efficacy (capture and sensing thresholds, lead impedance, handling characteristics, and clinical events). ThinLine coradial bipolar leads demonstrated lower capture thresholds than coaxial controls, comparable sensing, and acceptable impedances over 17,530 device months.