Key result
The new co-radial bipolar pacing leads demonstrated excellent or good X-ray visibility in 82 of 103 patients, with stable pacing characteristics and 0 lead-related complications over 18 months.
Why the study?
Do co-radial bipolar pacing leads demonstrate stable electrical characteristics and safety over 18 months in patients requiring pacing?
Population
103 patients requiring pacing leads who received 71 atrial and 89 ventricular leads.
Design
Cohort
Follow-up
18 months
Authors
Loading...
Supports short-term safety of co-radial bipolar leads; leaves open need for randomized trials versus conventional leads.
Observational (n=103)
Do co-radial bipolar pacing leads demonstrate stable electrical characteristics and safety over 18 months in patients requiring pacing?
Co-radial bipolar pacing leads demonstrate excellent handling, adequate visibility, and stable electrical performance up to 18 months.
Tang et al. (1997) reported an observational. Co-radial bipolar pacing leads (Intermedics Thin-Line) was evaluated on X-ray visibility, lead handling, impedance, sensitivity threshold, and pacing threshold. The new co-radial bipolar pacing leads demonstrated excellent or good X-ray visibility in 82 of 103 patients, with stable pacing characteristics and 0 lead-related complications over 18 months.